Skip to the interactive Type J converter

Instrumentation reference

J Type Thermocouple Chart

Convert Type J temperature and millivolts in either direction, account for a nonzero reference junction, and look up nominal ITS-90 checkpoints.

Direct range
−210°C to 1200°C
Nominal output
−8.095 to 69.553 mV
Reference
ITS-90, 0°C basis

Cold-junction-aware converter

Convert Type J Temperature and Millivolts

Uses the nominal Type J reference function and one canonical °C/mV calculation path.

At the default 100°C hot junction and 0°C reference junction, the nominal measured output is 5.268916 mV (displayed as 5.269 mV).

Nominal reference values

This conversion is not a probe accuracy, calibration, service-life, sheath-temperature, atmosphere-suitability, or project-approval result. Do not extrapolate beyond the stated function range.

Calculated checkpoints

Type J Thermocouple Reference Table

Limited nominal Type J ITS-90 checkpoints relative to a 0°C reference junction.

Search a temperature or EMF value. Use the converter above for intermediate temperatures or a nonzero reference junction.

Scroll horizontally to view additional columns and vertically to view additional rows. Column headers remain visible while scrolling.

Limited nominal Type J ITS-90 checkpoints calculated relative to a 0°C reference junction.
TemperatureNominal Type J output at 0°C referenceDerived and scope
Function segment
−210−346−8.095−809519.1−210 to 760°C segment
−200−328−7.890−789021.9−210 to 760°C segment
−150−238−6.500−650033.1−210 to 760°C segment
−100−148−4.633−463341.1−210 to 760°C segment
−50−58−2.431−243146.6−210 to 760°C segment
−40−40−1.961−196147.5−210 to 760°C segment
0320.000050.4−210 to 760°C segment
25771.277127751.8−210 to 760°C segment
501222.585258552.8−210 to 760°C segment
1002125.269526954.4−210 to 760°C segment
1503028.010801055.2−210 to 760°C segment
20039210.7791077955.5−210 to 760°C segment
25048213.5551355555.5−210 to 760°C segment
30057216.3271632755.4−210 to 760°C segment
35066219.0901909055.2−210 to 760°C segment
40075221.8482184855.2−210 to 760°C segment
45084224.6102461055.4−210 to 760°C segment
50093227.3932739356.0−210 to 760°C segment
550102230.2163021657.0−210 to 760°C segment
600111233.1023310258.5−210 to 760°C segment
650120236.0713607160.3−210 to 760°C segment
700129239.1323913262.2−210 to 760°C segment
750138242.2814228163.7−210 to 760°C segment
760140042.9194291963.9760°C segment boundary
800147245.4944549464.6760 to 1200°C segment
850156248.7154871564.0760 to 1200°C segment
900165251.8775187762.4760 to 1200°C segment
950174254.9565495660.7760 to 1200°C segment
1000183257.9535795359.3760 to 1200°C segment
1050192260.8906089058.3760 to 1200°C segment
1100201263.7926379257.8760 to 1200°C segment
1150210266.6796667957.6760 to 1200°C segment
1200219269.5536955357.2760 to 1200°C segment

This is a limited, independently calculated checkpoint set—not the complete NIST or ASTM table and not a probe tolerance or rating table. EMF values are nominal relative to 0°C.

How to Read and Use the Type J Chart

  1. Confirm that the sensor, extension wire, connector, and instrument are configured for Type J.
  2. For a true 0°C reference junction, read the row directly or enter the hot-junction temperature in the converter.
  3. For an ambient reference junction, enter its measured temperature. The converter subtracts its nominal Type J EMF from the hot-junction EMF.
  4. For reverse conversion, enter the measured differential EMF and the actual reference-junction temperature.
Measurement-chain equationEmeasured = EJ(Thot) − EJ(Treference)

To reverse the measurement, first calculate E0 = Emeasured + EJ(Treference), then solve the bounded Type J reference function.

Why the reference junction changes the reading

A thermocouple instrument measures a voltage difference, not the hot-junction voltage alone. A 25°C terminal block already corresponds to about 1.277 mV on the Type J reference function, so that contribution must be restored before converting the measured voltage back to temperature. Electronic cold-junction compensation performs this correction using a separate terminal-temperature sensor.

Reference Range, Tolerance Range, and Probe Rating

Reference function
−210°C to 1200°C defines the nominal mathematical temperature–EMF relationship used here.
Manufacturing tolerance
The BIPM guide lists IEC Type J Class 1 and Class 2 tolerance coverage from −40°C to 750°C. A tolerance is not the same as the reference-function range.
Actual probe rating
Wire diameter, sheath, insulation, atmosphere, exposure time, construction, and manufacturer limits determine usable service.

Do not treat 1200°C as a universal rating

ASTM’s public scope also distinguishes reference tables from recommended upper-temperature limits. Check the exact probe and installation before service near any limit.

Reference function is not measurement accuracy

Error sourceWhat to check
Wire tolerance and driftCalibration history, batch, exposure temperature, and time.
Cold-junction sensorTerminal temperature, gradients, and compensation accuracy.
Instrument and wiringType selection, polarity, input accuracy, extension wire, connectors, grounding, and noise.
InstallationImmersion, heat conduction, thermal contact, sheath condition, and gradients across damaged wire.
InhomogeneityBends, strain, corrosion, contamination, oxidation, and prior high-temperature exposure.

Worked Type J Thermocouple Examples

100°C with a 0°C reference

Assumption: nominal Type J behavior. EJ(100°C) = 5.268916 mV and EJ(0°C) = 0 mV, so the measured output is 5.268916 mV. The table displays 5.269 mV.

A real probe can differ because of wire tolerance and the complete measurement chain.

100°C with a 25°C reference

EJ(100°C) − EJ(25°C) = 5.268916 − 1.277288 = 3.991628 mV. Reading 3.991628 mV as if the reference were 0°C would be incorrect.

Reverse 3.991628 mV at 25°C

Add the reference contribution: 3.991628 + 1.277288 = 5.268916 mV. Bounded reverse conversion returns approximately 100.0°C.

The compensated EMF must remain within the published Type J inverse range.

Type J Materials, Service Limits, and Common Mistakes

Positive leg
Iron
Negative leg
Copper-nickel
Typical strength
Established industrial use and tolerance of reducing atmospheres when the probe construction is suitable.
Important weakness
The iron leg can rust in moist atmospheres, and irreversible errors can increase above about 500°C.

The BIPM guide notes magnetic transformation near 760°C and an iron crystal transformation near 910°C. It also warns that older Type J definitions differed: pairing modern sensors with old instrumentation can create significant error. Generic “constantan” formulations used in Types E, J, and T should not be assumed interchangeable.

Common Type J mistakes

  • Selecting Type K or another thermocouple curve in the instrument.
  • Reversing polarity or using incompatible extension wire and connectors.
  • Confusing mV and µV, creating a factor-of-1000 error.
  • Assuming the instrument terminals are at 0°C.
  • Extrapolating beyond the reference function or treating it as a calibration certificate.
  • Ignoring iron corrosion, bends, drift, and the probe’s high-temperature history.

Source, Scope, and Data-Rights Notes

Calculated from BIPM Type J ITS-90 functions; verified against NIST SRD 60 Version 3.0; source checked August 9, 2026.

The limited table is calculated locally from the published BIPM coefficients. NIST is used as an independent reference check, and ASTM is cited only for its public scope. No complete NIST or ASTM table is reproduced or offered as a download.

Dataset and Source-Check Details

Reference condition
Nominal Type J EMF relative to 0°C; ITS-90 temperature scale
Calculated fields
°F, mV, µV, local sensitivity, cold-junction compensation, and limited checkpoints
Internal precision
Full double precision; display rounded to 0.001 mV, nearest µV, 0.1°C/°F, and 0.1 µV/°C
Excluded material
Complete standards tables, full 1°C dataset, tolerances as a calculator, color codes, and product ratings
Download decision
No CSV; copy is limited to the current calculated result or displayed page values
Publisher
Turn2Engineering

J Type Thermocouple Chart FAQs

Approximately 5.269 mV nominal relative to a 0°C reference junction. A warmer reference junction reduces the measured differential voltage.

Use the Nominal Chart with the Real Measurement Chain

Use the table for quick 0°C-reference checkpoints and the converter for intermediate values or cold-junction compensation. For field decisions, verify thermocouple type, polarity, reference-junction measurement, probe construction, environment, instrument accuracy, and calibration history.

Scroll to Top